Steam Quality Testing
Thermo Diagnostics specializes in providing expert Steam Quality Testing services for Healthcare and Pharmaceutical facilities.
The Standards are:
ANSI/AAMI ST 79 – "Comprehensive guide to steam sterilization and sterility assurance in health care facilities"
EN 285 - Sterilization - Steam sterilizers - Large sterilizers
When is steam quality testing required?
According to the ST 79 Standard, testing requirements are "Event Based", not on a schedule:
• Upon installation or relocation of the sterilizer
• After any change to the steam distribution lines or boiler supply water
• When investigating sterilization process failures
• When periodically assessing sterilization loads for wet packs
For both Standards, the testing measures three steam attributes:
• Steam Dryness - Measures the amount of moisture content in the steam. (Causes wet packs or staining)
• Superheat - Measures the superheat. (Causes scorching and cracked plastic)
• Non-Condensable Gas Measures the amount of non- condensable gas, such as air. (Causes failed DART or BI tests)