    Steam Quality Testing

    Thermo Diagnostics specializes in providing expert Steam Quality Testing services for Healthcare and Pharmaceutical facilities.

     

    The Standards are:

    ANSI/AAMI ST 79 – "Comprehensive guide to steam sterilization and sterility assurance in health care facilities"

    EN 285 - Sterilization - Steam sterilizers - Large sterilizers

    When is steam quality testing required?

    According to the ST 79 Standard, testing requirements are "Event Based", not on a schedule:

    • Upon installation or relocation of the sterilizer

    • After any change to the steam distribution lines or boiler supply water

    • When investigating sterilization process failures

    • When periodically assessing sterilization loads for wet packs

     

    For both Standards, the testing measures three steam attributes:

    • Steam Dryness - Measures the amount of moisture content in the steam. (Causes wet packs or staining)

    • Superheat - Measures the superheat. (Causes scorching and cracked plastic)

    • Non-Condensable Gas Measures the amount of non- condensable gas, such as air. (Causes failed DART or BI tests)